VDI Crash on Start or VDI will not Start
Case:
VDI begins to start, and then VReadout crashes.
Similar Case:
VDI won't start after installing or updating Verisurf.
Description:
Verisurf 22.2.4 and later are set up to look for a 64-bit VDI. VDI version 3.46.12 and later are all 64-bit VDIs. Installing a version of Verisurf earlier than 22.2.4 may cause an issue with how the VDI is set up to start in Verisurf, requiring users to follow the steps below to resolve the issue.
Solution:
1. Install the 64-bit VDI version
verisurf.com/downloads
or
C:\Program Files\mastercam 20xx\mastercam
2. Upon completion of the installation, delete the following folder C:\Users\Public\Documents\VDI
Note: This will delete all of the VDI settings and will begin to rebuild as soon as Device Setup is selected in Verisurf
3. Launch Verisurf and select Device Setup
4. Choose the Device and Select Start
If additional assistance is still required after performing these steps, please call Verisurf Support at 1-888-713-7201
Related Articles
Temp Comp / Temperature Settings Crash
Case: When attempting to access temperature settings or inputting a material temp which results in a crash. Cause: The TemperatureCompensation.xml file is corrupt. Solution: 1.) Navigate to C:\Users\Public\Documents\VDI\Settings 2.) Delete ...
Quick Surface CRASH on startup prior to activation
Case: Quick Surface CRASH on startup, preventing the user from activating the software license Cause: The security policy on the application PC is preventing Quick Surface from reaching its license server. Solution: Disable the ethernet connection ...
4K and High Resolution Displays | Verisurf Graphics, Buttons, and Icons Too Small and Truncated Text including VDI
Case: On systems that are using high-resolution monitors (4K displays, 21:9 ultra-wide monitors), buttons, text, and images can appear to be truncated or shifted out of place from where they should be. Cause: The issue is not the resolution (number ...
Leica Absolute family of Laser Trackers and Measurement Parameters and Modes
Case: The sample rate is slow when measuring using Stationary Points. Similar Case: How is the measurement rate increased? The DRO seems to be lagging. Description: We often get asked about the sample rate, measurement parameters, and modes for the ...
The what, when, how, and why with Closed Face Reflector and Enabling a Vector Offset in Verisurf for Laser Trackers
Case: The what, when, how, and why with Closed Face Reflectors and Enabling a Vector Offset in Verisurf for Laser Trackers What: Closed Face Reflectors cause refraction (deflection) when passing through the Face of the Reflector, resulting in the ...